Hospital Supplies

I   Home    I    Equipment   I    Nurses   I    Medical Equipment   I    Dialysis Machine   I    Ultrasound   I    Dialysis   I    Nursing Agency   I    health Journal   I    First Aid   I    Cancer   I    Contact   I    Career   I


Thursday, November 10, 2011

Blood pressure monitor GE Dash 2500


The Dash® 2500 monitor from GE Healthcare is the latest vital signs monitor in the Dash Family

By combining Dash family performance and Gold Standard Dinamap Superstat NIBP technology, the Dash2500 fits for hospital supply in every departement. It allows you to deliver a new standard of clinical excellence to all patient groups in low-acuity settings such as ER, recovery/PACU, outpatient surgery, GI/endoscopy, progressive care, interventional radiology, general care and special care units.

The Dash 2500 monitor is a reliable, affordable bedside monitor that gives you the clinical intelligence you need to assess and treat your patients with speed, accuracy and precision. The Dash 2500 monitor includes sophisticated clinical parameters to capture vital patient measurements and exceptional cardiac monitoring to help accurately detect arrhythmias. The Dash 2500 monitor is designed for demanding clinical environments. Built with outstanding durability, it is also totally portable and easy to use.

The Dash 2500 includes:
• 3 and 5 lead ECG -> EK-Pro™ arrhythmia analysis
• Respiration monitoring
• Non-invasive blood pressure -> DINAMAP® SuperSTAT™, The Gold Standard for NIBP accuracy
• Pulse Oxymetry -> Masimo SET® or Nellcor® OxiMax® SpO2
• Temperature: Alaris® Turbo Temp®
• a large screen (10.4”)
• low weight only 5,5 Kilo’s

Product specifications


• Physical specifications
• Height 22.0 cm (8.7 in)
• Width 35.8 cm (14.1 in)
• Depth 17.0 cm (6.7 in)
• Weight 5.5 kg (12 lb)
• Environmental
• Operating temperature 5°C to 40°C (41°F to 104°F)
• Storage temperature -20°C to 60°C (-4°F to 140°F)
• Operating humidity 5 to 95% non-condensing
• Storage humidity 5 to 95% non-condensing
• Operating atmospheric 700 to 1060 hPa
• pressure
• Storage atmospheric 500 to 1060 hPa
• pressure
• Electrical
• AC input voltage 100 to 240 V
• AC input frequency 50/60 HZ
• AC input power 120 VA
• Internal battery 8.4 V nickel metal-hydride (NiMH)
• Power supply The Dash 2500 Patient Monitor
• can be powered from the internal
• battery or AC power
• Battery
• Capacity 8.4 V; 7.0 amp-hr
• Battery life Greater than 180 minutes using
• fully charged internal battery (NIBP:
• five min auto cycle with adult cuff.
• ECG, RESP, SpO2: Active. TEMP:
• predictive mode. Printer: printing
• two waveforms for one min every
• 20 min at 25 mm/sec)
• Charge time 4 hours maximum with the
• Monitor switched OFF
• 8 hours maximum with the
• Monitor switched ON

Heart Rate/Pulse

• Electrocardiography (ECG)
• Heart rate accuracy 30 to 300 bpm, ± 3 bpm or 3%
• of reading, whichever is greater
• Nellcor SpO2
• Range 20 to 250 bpm
• Accuracy and tolerance 20 to 250 bpm ± 3 digits
• Low perfusion 20 to 250 bpm ± 3 digits
• Masimo SpO2
• Range 25 to 240 bpm
• Accuracy and motion tolerance
• Without motion 25 to 240 bpm ± 3 digits
• With motion normal physiologic range,
• 25 to 240 bpm ± 5 digits
• Non-invasive blood pressure (NIBP)
• Adult/pediatric range 30 to 200 bpm
• Neonate range 30 to 220 bpm
• Accuracy ± 3.5% or 3bpm, whichever is greater

ECG

• Leads available
• 3-electrode I, II, III
• configuration
• 5-electrode I, II, III, aVR, aVL, aVF, and VA
• configuration
• Heart rate accuracy 30 to 300 bpm, ± 3 bpm or 3% of
• reading, whichever is greater
• Heart rate resolution 1 bpm
• Bandwidth 0.5 to 40 Hz +1/-6 dB
• 0.05 to 40 Hz +1/-6 dB
• 0.05 to 100 Hz +1/-6 dB
• Standardizing voltage 1 mV marker
• Common mode rejection 1 mV RTI or 10 mm p-p max
• displayed noise allowed with 20
• Vrms, 50-60 Hz input
• Input impedance
• Common mode > 2.5 MΩ at 10 Hz
• Differential > 2.5 MΩ from DC to 60 Hz
• 60 Hz tolerance up to 10 mV
• Pacemaker detection/rejection
• Input voltage range ± 2 to ± 700 mV
• Lead off sensing current < 0.1 μA DC signal leads,
• < 1 μA DC driven lead

Respiratory

• ECG-Derived respiration rate
• Leads available I or II
• Range 6 to 120 breaths/min (adult/pediatric)
• 6 to 180 breaths/min (neonate)
• Accuracy ± 2 breaths/min or ± 3% of reading,
• whichever is greater
• Resolution 1 breath/min
• Base impedance 100 to 2000 Ω
• Detection sensitivity 0.2 Ω at 30 breath/min with
• 500 Ω baseline impedance


The Dash® 2500 monitor from GE Healthcare is the latest vital signs monitor in the Dash Family
NIBP

• Method Oscillometric with step deflation
• Modes Manual, automatic, stat
• BP Measurement ranges
• Systolic 30 to 290 mmHg (adult/pediatric)
• 4.0 to 38.7 kPa (adult/pediatric)
• 30 to 140 mmHg (neonate)
• 4.0 to 18.7 kPa (neonate)
• MAP 20 to 260 mmHg (adult/pediatric)
• 2.7 to 34.7 kPa (adult/pediatric)
• 20 to 125 mmHg (neonate)
• 2.7 to 16.7 kPa (neonate)
• Diastolic 10 to 220 mmHg (adult/pediatric)
• 1.3 to 29.3 kPa (adult/pediatric)
• 10 to 110 mmHg (neonate)
• 1.3 to 14.7 kPa (neonate)
• Resolution 1 mmHg
• Accuracy Meets AAMI/ANSI standard
• SP10:2002
• Initial cuff 135 ± 15 mmHg default;
• inflation pressure user selectable (adult/pediatric)
• 100 ± 15 mmHg default;
• user selectable (neonate)
• Maximum determination 120s (adult/pediatric)
• time 85s (neonate)
• Over pressure monitor 300 to 330 mmHg (adult/pediatric)
• 150 to 165 mmHg (neonate)
• Hose/cuff interface Compatible with current Dash hoses
• Pulse rate When NIBP is the source, HR values
• are derived from the pulse rate that
• is determined by the oscillometric
• technique of measuring blood
• pressure. The rate source field is
• labeled NIBP.
• Adult/pediatric range 30 to 200 bpm (± 3.5% or 3 bpm)
• Neonate range 30 to 220 bpm (± 3.5% or 3 bpm)

Nellcor OxiMax SpO2

• Measurement range
• SpO2 1 to 100%
• Pulse rate 20 to 250 bpm
• Accuracy
• Saturation
• Adult 70 to 100% ±2 digits
• Neonate 70 to 100% ±3 digits
• Low perfusion 70 to 100% ±2 digits
• Pulse Rate
• Adult and neonate 20 to 250 bpm ±3 digits
• Low perfusion 20 to 250 bpm ±3 digits

Masimo SET SpO2

• Measurement range
• SpO2 1 to 100%
• Pulse rate 25 to 240 bpm
• Accuracy and motion tolerance
• Saturation
• Without motion– 70 to 100% ±2 digits
• adult/pediatric
• Without motion– 70 to 100% ±3 digits
• neonate
• With motion– 70 to 100% ±3 digits
• adult/pediatric/neo
• Low perfusion 70 to 100% ±2 digits,
• 0 to 69% unspecified
• Pulse Rate
• Without motion 25 to 240 bpm ±3 digits
• With motion normal physiologic range
• 25 to 240 bpm ±5 digits

Alaris Turbo Temp

• Scale ° Fahrenheit (F)
• ° Celsius (C)
• Predictive mode
• Range 35.6°C to 41.1°C (96.0°F to 106.0°F)
• Resolution 0.1°C (0.1°F)
• Monitor mode
• Range 26.7°C to 42.2°C (80.0°F to 108.0° F)
• Accuracy ± 0.1°C (± 0.2°F) (when tested in a
• calibrated liquid bath; meets ASTM
• E1112, Table 1, in range specified)
• Resolution 0.1°C (0.1°F)
• Probes Use only Alaris Turbo Temp probes
• and probe covers. The size, shape,
• and thermal characteristics of the
• probe covers can affect the performance
• of the instrument. Inaccurate
• readings or retention problems may
• occur unless Alaris Turbo Temp
• probes and probe covers are used.
• Determination time Approximately 10 seconds,


GE Health Care

P.O. BOX 900, FIN-00031 GE, Finland
Tel. + 3581039411. Fax + 35891463310

We have diet solution program for those people who wants to reduce body weight. just click here.


1 Comments:

  • This article was great in suggesting new brands and
    styles to try out.
    , I thought this article was suppose
    to help those of us who have never used Blood Pressure Monitor
    Thank you so much for this comprehensive list!
    I have read many reviews and done a ton of research
    . This has helped me make decisions on Blood Pressure Monitor.
    .I am completely happy with your website
    . All comments and articles
    are very useful and very good. Your blog is very
    careful-take control.
    I am loving all of the inside ,

    By Blogger Unknown, At June 26, 2014 at 10:20 PM  

Post a Comment

Subscribe to Post Comments [Atom]



<< Home